Description
Dimensions
- Height: 8.4″ (21.3 cm)
- Width: 10.3″ (26.2 cm)
- Depth: 10.3″ 26.2 cm)
- Weight: fully featured defibrillator/monitor (pacing and SpO2) 5.58 kg (12.3 lbs)
Power
- The device is an AC line operated device with an internal battery as back up.
- AC Power: 90-132 VAC 50/60 Hz, 198-264 VAC 50/60 Hz
- total power draw less than 75 Watts
- Internal battery back up: NiMh, batteries charge while device operates from AC power
- Operating time: a new fully charged internal backup battery will provide the following prior to shutdown: 120 minutes of monitoring, 135 minutes of monitoring without pulse oximeter, 90 360 joules discharges, and 70 minutes of monitoring plus pacing
- Typical battery charge time: <2 hours when the device is powered off and AC power is applied
- Low battery indication and message: when the device is unplugged from AC power, it switches to battery. When the battery gets low, the battery detection icon is indicated with low battery message in the status area, and warning tone occurs
Display
- Height: 3.4″ (8.63 cm)
- Width: 4.53″ (11.5 cm)
- Resolution: 230 x 240 dot color active LCD
- Display Time: a minimum of 4 seconds of ECG and alphanumeric for values, device instructions or prompts. Option to display one additional waveform.
- Sweep Speed: 25 mm/sec for ECG.
Data Management
The device can easily print a CODE SUMMARY report, including an introduction to the patient information and critical event record. The summary report also includes event and vital signs log, and waveform associated with certain events.
Communications
The device can print archived patient records and has two data communications port, which supports a serial data cable.
Monitor
- ECG: can be monitored through 3-wire or 5-wire ECG cables. Standard paddles or therapy electrodes are used for paddles lead monitoring
- Lead selections: Leads I, II and III with a 3-wire ECG cable, leads I, II, III, AVR, AVL, and AVF, V acquired simultaneously with a 5-wire ECG cable.
- ECG size: 4, 3, 2.5, 2, 1.5, 1, 0.5, 0.25 cm/mV
- Heart rate display: 20-350 bpm digital display
- Voice prompts: used for selected warnings and alarms, configured on/off
- SpO2: saturation range: 1 to 100%
- Saturation accuracy: (70-100%) (0-69% unspecified): Adults/ Pediatric: ±2 digits with no motion, ±3 digits with motion. Neonates: ±3 digits with or without motion
- SpO2 update averaging rate: user selectable 4, 8, 12 or 16 seconds
- SpO2 measurements: functional SpO2 values are displayed and stored
- Pulse rate range: 25 to 240 pulses per minute
- Pulse rate accuracy: ±3 digits without motion and ±5 digits with motion
Frequency Response
- Diagnostic: 0.05 to 150 Hz or 0.05 to 40 Hz (user configurable)
- Monitor: 0.67 to 40 Hz or 1 to 30 Hz (user configurable)
- Paddles: 2.5 to 30 Hz
- Analog ECG Output: 0.67 to 32 Hz (except 2.5 to 30 Hz for paddles ECG)
Reviews