l

Ut wisi enim ad minim veniam, quis laore nostrud exerci tation ulm hedi corper turet suscipit lobortis nisl ut

Recent Posts
2643-G and H Taft Avenue Pasay City Manila, PH 1300
26846 Oak Avenue Unit C Canyon Country, CA 91351
Image Alt

Covidien Valleylab Force EZ

Instant Response technology provides surgeons with improved performance or lower power settings, minimizing the risk of tissue damage and neuromuscular stimulation, and decreasing the need to “turn up the generator.” The Force EZ generator adjusts automatically, responding to tissue changes, maintaining power delivery, and minimizing drag. Capacitive coupling is reduced by 30-50% when using Instant Response technology. This reduction is achieved by limiting the RMS voltage and the high-frequency harmonics. Lower voltage means less neuromuscular stimulation and a more precise delivery of energy to reduce collateral damage.

Instant Response technology features an advanced feedback system that recognizes changes in tissue 200 times per second and adjusts voltage and current accordingly to maintain appropriate power. This unique capability differentiates Instant Response generators by their high power efficiency rating (PER). That’s smart generator technology. PER indicates the ability of a generator to accurately deliver the selected power into a wide range of tissue types. With a PER of 97, the Force EZ generator provides a consistent surgical effect.

  • Instant Response technology ensures that the power delivered remains virtually constant, regardless of the tissue type
  • Improved performance at lower power settings minimizes the risk of tissue damage and neuromuscular stimulation
  • Three internal microcontrollers reduce system reaction time and increase the system’s processing speed
  • Easy to use, cost effective system when performance is required and price is of equal concern
  • Two cut modes, both controlled by Instant Response technology, offer surgeons a variety of choices:
    • Pure Cut for a clean, precise cut
    • Blend for cutting with hemostatsis
  • Two primary coag modes:
    • Low for lower voltage contact coagulation suitable in laparoscopic and delicate tissue work
    • High for efficient noncontact coagulation in most applications

Description

Dimensions

  • Height: 5 inches (12.7 cm)
  • Width: 16 inches (40.6 cm)
  • Length: 15.6 inches (39.5 cm)
  • Weight: < 15 lbs (6.8 kg)

Output Waveforms

  • Bipolar: 470 kHz sinusoid
  • Monopolar Cut:
    • Pure: 393 kHz sinusoid
    • Blend: 393 kHz bursts of sinusoid, recurring at 27 kHz intervals. 50% duty cycle envelope
  • Monopolar Coag:
    • Desiccation:
      • Low 1: 240 ± 40 kHz sinusoid recurring at 39 kHz. 8% duty cycle
      • Low 2: 393 kHz sinusoid
      • Low 3: 393 kHz sinusoid
    • Fulguration:
      • High 1: 470 ± 40 kHz damped sinusoidal bursts with a repetition frequency of 57 kHz
      • High 2: 470 ± 40 kHz damped sinusoidal bursts with a repetition frequency of 30 kHz
  • Output power changes by less than 15% or 5 watts, whichever is greater, as the line voltage varies from 90-135 volts (into a 300 ohm load)

Low Frequency Leakage (50-60 Hz)

  • Source current, patient leads, all outputs tied together
  • Normal polarity, intact chassis ground: < 10µA
  • Normal polarity, ground open: < 50µA
  • Reverse polarity, ground open: < 50µA
  • Sink current, 140V applied, all inputs: < 50µA

High Frequency Leakage

  • Bipolar: Less than 60 mArms
  • Monopolar: Less than 150 mArms

Operating Parameters

  • Ambient Temperature range: 10° to 40° C (50° to 104° F)
  • Relative humidity: 30% to 75%, non-condensing
  • atmospheric Pressure: 700 to 1060millibars
  • Warm-up Time: If you transported or stored the generator at temperatures outside the operating temperature range, allow one hour for the generator to reach room temperature before use

Transport and Storage

  • Ambient Temperaturen range: 34° to 65° C (-29° to 149° F)
  • Relative Humidity: 25% to 85%, non-condensing
  • Atmospheric Pressure: 500 to 1060 millibars
  • Duration of Storage: Force EZ-C Series Service Manual Transport and StorageAmbient temperature-34° to 65° C (-29° to 149° F)rangeRelative humidity25% to 85%, noncondensingAtmospheric pressure500 to 1060 millibarsDuration of storageIf you stored the generator longer than one year,replace the battery for battery-backed RAM andcomplete a full test and recalibration before use.

Input Power Requirements

  • 110-120Vac:
    • Operating Range: 85-140Vac
    • Max Current Cut: 8A
    • Max Current Coag: 4.2A
  • 220-240Vac:
    • Operating Range: 170-280Vac
    • Max Current Cut: 4A
    • Max Current Coag: 2.1A

Reviews

There are no reviews yet.

Be the first to review “Covidien Valleylab Force EZ”