Description
Ventilator Dimensions
- Height: 11 in
 - Width: 13.5 in
 - Depth: 10 in
 
Graphic Display Dimensions
- Height: 9.5 in
 - Width: 13 in
 - Depth: 2.5 in
 - Weight: 5.5 lbs
 
Pedestal Stand Dimensions
- Height: 40.5 in
 - Width: 24 in
 - Weight: 9 lbs.
 
Compressor Dimensions
- Height: 36 in
 - Width: 22 in
 - Depth: 21.5 in
 - Weight: 11o lbs
 
Environmental
- Temperature
- Storage and Shipping -40 to 158 ºF (-40 to 70 ºC)
 - Checkout and Operating 50 to 104 ºF (10 to 40 ºC)
 
 - Altitude
- Checkout, Operating, Transport and Storage 0 to 9,000 feet (14.7 to 10.5 PSIA/760 to 543 mm Hg)
 
 - Humidity
- Storage and Shipping 0 to 99% Relative Humidity — Non-condensing
 - Checkout and Operating 0 to 95% Relative Humidity — Non-condensing
 
 
Controls
- Mode: Flow Cycle AC, AC, SIMV/IMV, SIMV/PSV, Flow Cycled SIMV CPAP/PSV
 - Rate: 1 to 150 bpm
 - Inspiratory Time: 0.1 to 3.0 sec
 - PEEP: 0 to 30 cmH2O
 - Inspiratory Pressure: 0 to 72 cmH2O
 - Manual Breath: x1
 - Assist Sensitivity: 0.02 to 5.0 L/min
 - Over Pressure Relief: 15 to 75 cmH2O
 - % O2: 21 to 100%
 - Apnea Alarm (selectable): 5, 10, 20 or 30 sec
 
Monitors
- Breath Rate: 0 to 255 bpm
 - Patient Initiated Indicator: LED
 - Minute Volume: 0 to 30.0 L/min
 - Tidal Volume: 0 to 500 ml (Inspired and Expired)
 - %Tube Lea:k 0 to 100%
 - Inspiratory Time: 0.1 to 3.10 sec
 - Expiratory Time: 0 to 99.9 sec
 - I:E Ratio: 9.1 to 1:9.9
 - Peak Inspiratory Pressure: 0 to 99 cmH2O
 - Mean Airway Pressure: 0 to 75.0 cmH2O
 - PEEP: 0 to 30.0 cmH2O
 - Air Pressure: 0 to 100 psig
 - O2 Pressure: 0 to 100 psig
 - Proximal Airway Pressure: 10 to 100 cmH2O (Gauge)
 - Hourmeter: 0 to 99,999 hours
 - Test: Push button
 
Alarms / Limits
- High Breath Rate: 3 to 255 bpm
 - Low PEEP/CPAP: -5 to 30 cmH2O
 - Low Inspiratory Pressure: 0.25 (High Pressure Limit – Low PEEP/CPAP) + Low PEEP/CPAP
 - Patient Circuit: LED On/Off
 - Failed to Cycle: On/Off
 - Low Gas Supply: LED On/Off
 - Apnea: LED On/Off
 - Pressure Settings Incompatible: LED On/Off
 - Prolonged Inspiratory Pressure: LED On/Off
 - Flow Sensor: LED On/Off
 - Low Battery: LED On/Off
 - Alarm Silence: 60 sec
 - Visual Reset: Push Button
 - High Pressure Limit: 10 to 75 cmH2O
 - Line Power: Green/Red LED
 
Electrical
- Voltage
- 100 V – 80 to 110 VAC
 - 120V – 96 to 132 VAC
 - 220V – 176 to 242 VAC
 - 240V – 192 to 264 VAC
 
 - Current
- 100V – 1.0 A maximum
 - 120V – 1.0 A maximum
 - 220V – 0.5 A maximum
 - 240V – 0.5 A maximum
 
 - Frequency
- 100V – 50.60 Hz
 - 120V – 50/60 Hz
 - 220V – 50/60 Hz
 - 240V – 50/60 Hz
 
 - Fuses
- 100/120V – T o.5 A, 5 x 20 mm
 - 230/240V – T 0.5 A, 5 x 20 mm
 - (For 220V operation, set the power entry module seclector switch to 230V setting)
 
 
					
					
					
        
		