The Physio-Control Lifepak 9 is a microprocessor-based defibrillator, and is available in three models: 9, 9P (pacing), and 9A. These units have a built-in recorder and sealed lead acid battery back-up. Lifepak 9s meet current ACLS standards and allows for 17 energy selections, HR alarms, 5-lead monitoring, recorder and CODE SUMMARY critical event record. The Lifepak 9 is defibrillator/monitor, cardioscope, and non-invasive pacemaker. It provides 17 energy selections, heart rate alarms, three-lead ECG monitoring, recorder and Code Summary™ critical event recorder. The microprocessor-based Lifepak 9 includes an internal battery supply and is available with optional pediatric paddles, internal paddles, pacing cable and pads, and ECG patient cable and leads. Isolated ECG via Quick-Look defibrillator paddles, Fast-Patch disposable defibrillation/ECG electrodes or three-lead patient cable.

  • Defibrillator with ECG and recorder
  • The Physio-Control LIFEPAK 9 defibrillator/monitor was designed to meet the needs of clinical, biomedical engineering, and purchasing professionals throughout the world
  • This unit issimple 1- 2-3 operation and user-friendly display, plus its compact, portable, space-saving design, revolutionizes in-hospital defibrillation
  • ECG Lead Selections: STD, Paddles, I, II, III
  • Input: Isolated ECG via Quick-Look defibrillator paddles, Fast-Patch disposable defibrillation/ECG electrodes or three-lead patient cable
  • Patient Cable Length: 13-foot total length (10-foot w/ three-foot leads)
  • Common Mode Rejection: 100dB min. with respect to chassis ground and 65dB min. with respect isolated ground, when measured at 60Hz (common mode range for patient cable input is 10 Volts peak w/ respect to isolated ground)
  • Monitor Display Size: 4” (w) x 3” (h)
  • Sweep Speed: 25 mm/sec
Category:

Description

Dimensions

  • Height: 35.2 cm (13.9 in)
  • Wdith: 29.7 cm (11.7 in)
  • Depth: 31cm (12.2 in)
  • Weight: 12.7 kg ( 28 lbs)

Generial specifications

  • Power Cord Length: 3m (10 Ft)
  • Standard Paddle Electrode area: 82 square cm
  • AC Input Option: 120 or 240 Vac nominal line voltage without adjustment. 50 or 60Hz, with adjustment in SET UP menu.
  • Battery type: Sealed lead-acid, 16 Vdc nominal, 3 Amp-hour
  • low battery indicator: Advises operator toconnect AC power.
  • Battery Charge Indicator: Illuminates when battery is charging.
  • Battery Charge Time: 24 hours to full capacity.
  • Service Indicator: Indicates self-diagnostic routines have detected improper operation requiring service attention.
  • Power Consumption: 160 Watts maximum while monitoring with recorder ON and defibrillator charging.

ECG Monitor

  • ECG Lead Selections: Std, Paddles, I, II, III
  • Input: Isolated ECG via QUIK-LOOK® defibrillator paddles, FAST-PATCH® disposable defibrillation/ ECG electrodes or 3-lead patient cable.
  • Patient Cable Length: 4.0m (13 ft) total length; 3.1m cable (10 ft) with 0.9m leads (3 ft)
  • Common Mode Rejection: 100dB minimum with respect to chassis ground and 65dB minimum with respect to isolated ground, when measured at 60Hz. Common mode range for patient cable input ≥ 10 Volts peak with respect to isolated ground.
  • Monitor Display Size:102mm (4 in) wide x 76mm (3 in) tall, non-fade
  • Sweep Speed: 25mm/second
  • Frequency Response: Non-diagnostic
  • ECG Leads:1.0 to 40Hz (-3dB)
  • Paddles: 2.2 to 20Hz (-3dB)
  • ECG Size: Adjusts amplitude of ECG trace on monitor and strip chart recorder.
  • Heart Rate Meter: 3 digit readout displays rates from 20 to 300 bpm. Heart rates outside this range do not yield valid systole tones or heart rate display. Heart Rate Alarm: User selectable alarm limits. Three high/low settings (150/40, 120/60, 160/90). Other limit options available via SET UP menu.
  • 1 mV Cal: Momentary pushbutton switch on rear panel simulates a 1 mV signal pulse to the ECG input.
  • ECG Output:1 Volt/mV

Defibrillator

  • Defibrillator Waveform: 5 millisecond monophasic pulse (Edmark)
  • Energy Select:
    • External Paddles: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 20, 30, 50, 100, 200, 300, 360 Joules
    • Internal Paddles: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 20, 30, 50 Joules
  • Charge Controls: Independent momentary push button controls on front panel and APEX paddle.
  • Paddle Cord Length:3m (10 ft)
  • Charge Indicators: Flashing lamps on paddle and front panel pushbutton along with increasing stored energy display indicate charge in progress. Upon full charge, energy available is displayed on monitor and charge completed tone sounds.
  • Charge Time: Charge to 360 Joules in less than 10 seconds with a fully charged battery.
  • SYNC: Synchronizes defibrillator pulse to patientgenerated QRS complex.
  • SYNC Indicator: Inverted triangle marker on displayed ECG waveform identifies synchronizer trigger point with respect to patient’s QRS complex.

Thermal Array Recorder

  • Paper Size: 50mm wide x 30m (100 ft)
  • Paper Speed: 25mm/second
  • Delay: ECG prints 8 seconds after first appearing on the monitor.
  • Frequency Response: Non-diagnostic. Diagnostic available via SET UP menu.
  • ECG Leads (non-diagnostic): 1.0 to 40Hz (-3dB)
  • ECG Leads (diagnostic): 0.05 to 100Hz (-3dB)
  • Paddles: 2.2 to 20Hz (-3dB)
  • Annotation: Time, date, ECG lead, ECG gain, heart rate, defibrillation parameters, test load discharges
  • CODE SUMMARY Critical Event Record: Digitally stored record of ECG and device parameters

Environmental

  • Atmospheric Pressure: 797mmHg to 500mmHg (-570 to 11000 ft)
  • Relative Humidity:
    • 0 to 95% (non-condensing) at 0 to 34°C
    • 0 to 80% (non-condensing) at 35 to 45°C
  • Temperature Range: 0 to 45°C operating -30 to 65°C storage
  • All specifications at 20°C unless otherwise stated.