Description
Dimensions
Incubator with Elevating Base:
- Height: Overall height 54 to 62 inches (136.2 to 157.5 cm)
- Elevating base adjustable from 35 to 43 inches (88.9 to 109.2 cm) mattress to floor
- Depth: 71.1 cm/28 in
- Width: 83.8 cm/33 in
- Weight: 89 kg/195 lb
Incubator with Elevating Base:
- Height: Overall height 54 to 62 inches (136.2 to 157.5 cm)
- Elevating base adjustable from 35 to 43 inches (88.9 to 109.2 cm) mattress to floor
- Depth: 71.1 cm/28 in
- Width: 83.8 cm/33 in
- Weight: 89 kg/195 lb
Drawer Packages for Elevating Base or Stationary Pedestal:
- Overall: 28 x 29 x 46 cm/11 x 11.5 x 18.25 in
- Single deep drawer: 25 x 23 x 42 cm/10 x 9 x 16.5 in
- Dual drawer: 11 x 23 x 42 cm/4.5 x 9 x 16.5 in
Incubator with Cabinet:
- Height: 141 cm/55 in
- Depth: 64.6 cm/25.5 in
- Width: 88.9 cm/35 in
- Weight: 84 kg/185 lb
Incubator Characteristics:
- Casters: 5 inch diameter, 2 locking, 2 non-locking
- Mattress: 34.8 x 65 cm/13.7 x 25.6 in
- Mattress to Hood: 16 in/41 cm
- Door Height: 11 in/28 cm
- Tilt: 0°, 3° and 6° (CP 3000)
- Positions: Continuous from 0° to 12° (CP 4000)
- Port holes: 6
- Tubing ports: 6
Electrical Power Requirements
- 120 Vac 60Hz
- Models: 115 V ±10% 5.7 amps
- Conforms to IEC 601.1, UL 544 and CSA 22.2 No. 125 requirements
- Nominal power consumption: 450 watts at maximum heater output
- Line voltage compensation: Input voltage is monitored and the heat output is adjusted to compensate for line voltage variations
- Leakage currents: Less than 100 micro amps on 120V units
- Leakage at patient probe: Less than 50 micro amps on 120V unit
Operation Enviroment
- Operating temperature range: 20° to 30°C/68° to 86°F
- Storage temperature range: -25° to 60°C/13° to 140°F
- Operating humidity range: 0 to 90%
System Control Characteristics
- Microprocessor based control system: Self-test functions are performed at power-up and during normal operation
- Patient Control (Servo) Mode: 35° to 37°C/95° to 98.6°F
- Air Control (Manual) Mode: 20° to 37°C/73.4° to 98.6°F
- Up to 39°C/102.2°F with control panel override switch
- Air Velocity: Average <10 cm/sec with inner wall
- Temperature variability: Less than ±0.3°C
- Resolution: ±0.1°C or 0°F
- Probe interchangeability: ±0.1°C/0.2°F
- Average oxygen input concentration range:
- 5 L/min 25 to 45%
- 10 L/min 35 to 65%
- 15 L/min 45 to 70%
- Humidification: Standard – 10 to 75% dependant on nursery environment and incubator temperature setting
- Servo – 40
Reviews
There are no reviews yet.