l

Ut wisi enim ad minim veniam, quis laore nostrud exerci tation ulm hedi corper turet suscipit lobortis nisl ut

Recent Posts
2643-G and H Taft Avenue Pasay City Manila, PH 1300
26846 Oak Avenue Unit C Canyon Country, CA 91351
Image Alt

BiPAP Vision

Integrated Display Screen

Integrated Display Screen displays easy-to-read real time graphics in waveform or bar scale format, providing the following measured and calculated numerical values to assist with ventilator management:

  • IPAP
  • Total Breath Rate
  • Exhaled Minute Ventilation
  • % Patient Triggered Breaths
  • Exhaled Tidal Volume
  • EPAP
  • Ti/Tot
  • Peak Inspiratory Pressure
  • Patient or Total Leak

Modes

  • CPAP and S/T are available for maintaining airway patency and providing pressure support ventilation.
  • PAV * /T (Proportional Assist Ventilation/Timed mode) helps to promote improved patient comfort and enhanced patient/ventilator synchrony through independent adjustment of Volume and Flow assist based on the patients disease state.

Oxygen Control up to 100%

Allows you to blend oxygen within the unit in concentrations ranging between 21-100% without auxiliary oxygen tubing attached to the mask.

Pressure Regulation at the Mask

Pressure regulation at the mask allows you to regulate pressure therapy to assure delivery where it counts (not inside the machine), and automatically adjusts for variations.

Upgradeable Design Provides Flexibility

Upgradeable features provide flexibility in your purchase decision, now and in the future. The Vision System gives you the ability to upgrade instead of buying a new machine for each medical revolution.

BiPAP® Auto-Trak Sensitivity™

Optimum Response Automatically. The innovative development ofAuto-TrakSensitivityTM enables the BiPAP Vision System to automatically respond to changing breathing patterns and circuit leaks by identifying leaks and adjusting variable sensitivity thresholds for optimum response with every breath.
Category:

Description

Control Ranges and Increments: Parameters
Parameters Control Range Control Increment
IPAP 4 to 40 cm H2O 1 cm H2O
EPAP 4 to 20 cm H2O 1 cm H2O
CPAP 4 to 20 cm H2O 1 cm H2O
Rate 4 to 40 BPM 1 BPM
Timed Inspiration 0.5 to 3.0 sec 0.1 sec
IPAP >Rise Time 0.05 to 0.40 sec 4 set points: .05, .1, .2, .4 sec
Oxygen Concentration (%O2)* 21 to 100% 4% from 21 to 25%
5% from 25 to 100%

*With optional Oxygen Module

Control Ranges and Increments: Alarms

Alarm Control Range Control Increment
High Pressure 5 to 50 cm H2O 1 cm H2O
Low Pressure Disabled;0 to 40 cm H2O 1 cm H2O
Low Pressure Delay 0 to 60 sec 1 sec
Apnea Disabled; 20 to 40 sec 4 set points: Disabled, 20, 30, 40 sec
Low Minute Ventilation Disabled; 1 to 99 L/min 1 L/min
High Rate 4 to 120 BPM 1 BPM
Low Rate 4 to 120 BPM 1 BPM

Display Ranges and Increments

Parameters Control Range Control Increment
IPAP 0 to 50 cm H2O 1 cm H2O
EPAP 0 to 50 cm H2O 1 cm H2O
CPAP 0 to 50 cm H2O 1 cm H2O
Rate 0 to 150 BPM 1 BPM
Exhaled Tidal Volume (VT) 0 to 4000 ml 1 ml
Minute Ventilation (Min Vent) 0 to 99 L/min 1 L/min
Total Leak (Tot Leak) 0 to 300 L/min 1 L/min
Patient Leak(Pt. Leak) 0 to 300 L/min 1 L/min
Peak Inspiratory Pressure (PIP) 0 to 50 cm H2O 1 cm H2O
Percent of Patient Triggered Breaths (Pt. Trig) 0 to 100% 1%
Ti/Ttot 0 to 100% 1%