Bear Cub 750 PSV

The BEAR CUB™ 750 PVS ventilator can be classified as a microprocessor controlled, time cycled, pressure limited, dual flow, neonatal/pediatric critical care ventilator. Breaths are initiated either by the ventilator (controlled), by patient activity (assisted), or by the operator (manual).

  • Pneumatic Ventilator
  • Gas Inlet Conditioning
  • Accurate blending of gases in the O2 blender
  • Flow Control Valve
  • Auxiliary Gas Outlet
  • Pressure Control System
  • Flow Sensor
Category:

Description

Ventilator Dimensions

  • Height: 11 in
  • Width: 13.5 in
  • Depth: 10 in

Graphic Display Dimensions

  • Height: 9.5 in
  • Width: 13 in
  • Depth: 2.5 in
  • Weight: 5.5 lbs

Pedestal Stand Dimensions

  • Height: 40.5 in
  • Width: 24 in
  • Weight: 9 lbs.

Compressor Dimensions

  • Height: 36 in
  • Width: 22 in
  • Depth: 21.5 in
  • Weight: 11o lbs

Environmental

  • Temperature
    • Storage and Shipping -40 to 158 ºF (-40 to 70 ºC)
    • Checkout and Operating 50 to 104 ºF (10 to 40 ºC)
  • Altitude
    • Checkout, Operating, Transport and Storage 0 to 9,000 feet (14.7 to 10.5 PSIA/760 to 543 mm Hg)
  • Humidity
    • Storage and Shipping 0 to 99% Relative Humidity — Non-condensing
    • Checkout and Operating 0 to 95% Relative Humidity — Non-condensing

Controls

  • Mode: Flow Cycle AC, AC, SIMV/IMV, SIMV/PSV, Flow Cycled SIMV CPAP/PSV
  • Rate: 1 to 150 bpm
  • Inspiratory Time: 0.1 to 3.0 sec
  • PEEP: 0 to 30 cmH2O
  • Inspiratory Pressure: 0 to 72 cmH2O
  • Manual Breath: x1
  • Assist Sensitivity: 0.02 to 5.0 L/min
  • Over Pressure Relief: 15 to 75 cmH2O
  • % O2: 21 to 100%
  • Apnea Alarm (selectable): 5, 10, 20 or 30 sec

Monitors

  • Breath Rate: 0 to 255 bpm
  • Patient Initiated Indicator: LED
  • Minute Volume: 0 to 30.0 L/min
  • Tidal Volume: 0 to 500 ml (Inspired and Expired)
  • %Tube Lea:k 0 to 100%
  • Inspiratory Time: 0.1 to 3.10 sec
  • Expiratory Time: 0 to 99.9 sec
  • I:E Ratio: 9.1 to 1:9.9
  • Peak Inspiratory Pressure: 0 to 99 cmH2O
  • Mean Airway Pressure: 0 to 75.0 cmH2O
  • PEEP: 0 to 30.0 cmH2O
  • Air Pressure: 0 to 100 psig
  • O2 Pressure: 0 to 100 psig
  • Proximal Airway Pressure: 10 to 100 cmH2O (Gauge)
  • Hourmeter: 0 to 99,999 hours
  • Test: Push button

Alarms / Limits

  • High Breath Rate: 3 to 255 bpm
  • Low PEEP/CPAP: -5 to 30 cmH2O
  • Low Inspiratory Pressure: 0.25 (High Pressure Limit – Low PEEP/CPAP) + Low PEEP/CPAP
  • Patient Circuit: LED On/Off
  • Failed to Cycle: On/Off
  • Low Gas Supply: LED On/Off
  • Apnea: LED On/Off
  • Pressure Settings Incompatible: LED On/Off
  • Prolonged Inspiratory Pressure: LED On/Off
  • Flow Sensor: LED On/Off
  • Low Battery: LED On/Off
  • Alarm Silence: 60 sec
  • Visual Reset: Push Button
  • High Pressure Limit: 10 to 75 cmH2O
  • Line Power: Green/Red LED

Electrical

  • Voltage
    • 100 V – 80 to 110 VAC
    • 120V – 96 to 132 VAC
    • 220V – 176 to 242 VAC
    • 240V – 192 to 264 VAC
  • Current
    • 100V – 1.0 A maximum
    • 120V – 1.0 A maximum
    • 220V – 0.5 A maximum
    • 240V – 0.5 A maximum
  • Frequency
    • 100V – 50.60 Hz
    • 120V – 50/60 Hz
    • 220V – 50/60 Hz
    • 240V – 50/60 Hz
  • Fuses
    • 100/120V – T o.5 A, 5 x 20 mm
    • 230/240V – T 0.5 A, 5 x 20 mm
    • (For 220V operation, set the power entry module seclector switch to 230V setting)